Table 2 Averaged short time correlations

From: Technical and clinical analysis of microEEG: a miniature wireless EEG device designed to record high-quality EEG in the emergency department

Notch filter

Overall

Eyes open

Eyes closed

Off

0.860 ± 0.106

0.849 ± 0.109

0.877 ± 0.099

On

0.911 ± 0.070

0.902 ± 0.076

0.926 ± 0.058

  1. Correlations between themicroEEG and standard EEG computed for 1-s windows and averaged over the recording for all electrodes. Filters: 0.5-70 Hz bandpass.