Skip to main content

Table 3 Time and frequency domain properties

From: Technical and clinical analysis of microEEG: a miniature wireless EEG device designed to record high-quality EEG in the emergency department

 

Time domain

Spectral

 

Stdev(μV)

Kurt

Mob (Hz)

Comp

MF (Hz)

SEF75 (Hz)

SE

mEEG (Notch off)

67.9

9.1

108.5

2.9

9.3

9.4

11.8

STD (Notch off)

74.7

6.9

149.4

2.3

15.6

29.0

11.4

mEEG (Notch on)

65.6

10.0

69.0

3.5

6.0

7.9

11.9

STD (Notch on)

67.9

8.9

71.8

3.2

6.8

9.4

12.0

  1. microEEG and standard EEG recorded in parallel. Filters 0.5-70-Hz bandpass. Data shown are averages over all channels. mEEG, microEEG; STD, standard; Stdev, standard deviation; Kurt, kurtosis; Mob, mobility; Comp, complexity; MF, mean frequency; SE, spectral entropy. Kurtosis, mobility, complexity, and spectral entropy are dimensionless.