Accuracy of Broselow tape in estimating the weight of the child for management of pediatric emergencies in Nepalese population

Abstract

Background

Children with emergency conditions require immediate life-saving intervention and resuscitation. Unlike adults, the pediatric emergency drug dose, equipment sizes, and defibrillation energy doses are calculated based on the weight of the individual child. Broselow tape is a color-coded length-based tape that utilizes height/weight correlations for children. However, in low-income countries like Nepal, due to factors like undernutrition, the Broselow tape may not accurately estimate weight in all ranges of pediatric age group.

Methods

This study was conducted in the Department of Pediatrics of Dhulikhel Hospital, Kathmandu University Teaching Hospital, in children less than 15 years of age. Our study aims to prospectively compare the actual weights of urban and rural Nepalese children with the estimated weights using the Broselow tape (2017 edition) and the updated APLS formula. The errors in the selection of endotracheal tube size and adrenaline dose using the Broselow tape were also explored.

Results

This study included 315 children with male to female ratio of 0.63:1. They were divided into 3 groups according to their estimated weight by the Broselow tape into < 10 kg, 10–18, and > 18 kg. There was a total agreement of the estimated color zone according to the Broselow tape with the actual weight in the gray zone (p = 0.01). There was a positive relationship between the actual body weight and the estimated body weight (correlation (r = 0.970, p = 0.01) and accuracy (r2 = 0.941)). Our analysis showed that the accuracy of estimated weight with the Broselow tape decreases with increasing weight of children. The precision of the tape was relatively high in the lower length zones as compared to the higher length zones. The estimated size of the endotracheal tube (p = 0.01) and adrenaline dose (p = 0.08) by the Broselow tape was in agreement with that estimated using PALS formula in weight group of less than 18 kg, but decreases as the estimated weight increases further.

Conclusions

The accuracy of the Broselow tape in estimating the weight of a child, endotracheal tube size, and dose of adrenaline is higher in weight group of less than 18 kg, and accuracy decreases as the weight of child increases. The Broselow tape should be avoided in children weighing more than 18 kg. Hence, PALS age-based formula for ET tube size estimation and weight-based formula for adrenaline dose calculation are recommended for children weighing more than 18 kg.

Background

Children with a wide variety of urgent medical and surgical conditions visit the emergency department (ED), which requires immediate life-saving intervention and resuscitation. The accurate measurement or estimation of the weight of a child is crucial for the effective and optimal acute management of pediatric emergencies. In contrast to adults, the pediatric emergency drug dose, equipment sizes, and defibrillation energy doses are calculated based on the weight of the individual child and are a challenge for the treating emergency physician. The medical error related to the calculation of pediatric medication dosage is very high [1]. Incorrect estimation of patient weight, leading to incorrect drug dosing, is one of the most frequently reported errors [2]. It is not always feasible to measure the weight of a child using the standard weighing machine in the ED where the condition is critical and immediate action is required. When the accurate weight of the child cannot be obtained, it is usually calculated using an age-based formula such as advanced pediatric life support (APLS) formula which can be incorrect and time-consuming [3]. Wrong estimation of weight or incorrect calculation of drug dosage or equipment size could result in grave consequences.

Length-based weight estimation is developed as a different alternative to estimate the weight. The Broselow pediatric emergency tape is a color-coded length-based tape measure that was developed using height/weight correlations for children who have a maximum weight of roughly 36 kg from a nationally representative sample of children in the USA [4, 5]. The Broselow tape also provides medical instructions including the medication dosages, the size of the equipment, and the level of shock voltage when using a defibrillator. It is recognized in most medical textbooks and publications as a standard for the emergency treatment of children and is recommended by the Advanced Trauma Life Support and Pediatric Advanced Life Support [4, 6]. In a critical life-threatening condition, it is not judicious to consume the valuable time needed to evaluate, initiate, and monitor patient treatment to calculate the estimated weight, the equipment sizes, and the drug doses. The pre-calculated dosing in the Broselow tape facilitates rapid weight estimation, saves time by providing corresponding drug dosing, and alleviates stress during pediatric resuscitation [5, 7,8,9]. In a simulated pediatric emergency, color coding significantly reduced the deviation from recommended doses [8]. The studies done in various parts of the world have shown that the Broselow tape accurately measured the weight of children [10,11,12] while some studies showed it to be inaccurate [13,14,15,16,17]. The scenario is quite different in low-income countries like Nepal [15, 18, 19]. The undernutrition is still a major concern among pediatrics population in Nepal [20]. So, the recent modifications made given of obesity prevalence in western society do not address the issues of developing countries, which may lead to substantial overestimation of the weight of the children and potentially dangerous drug dosing and equipment selection [21]. A study done in urban Nepal by Shrestha et al. showed that the Broselow tape (2007 B edition) had only moderate accuracy for weight estimation [22].

This study aimed to prospectively compare the actual weights of urban and rural Nepali children with the estimated weights using the Broselow tape (2017 edition) and the updated APLS formula. The errors in the selection of endotracheal tube (ET) size and adrenaline dose using the Broselow tape were also explored.

Methods

Study design

This was a prospective cross-sectional study conducted from March to June 2019.

Study setting

The study was conducted in the Department of Pediatrics (Outpatient and Immunization Section) of Dhulikhel Hospital, Kathmandu University Teaching Hospital, Kavrepalanchowk. The hospital is a tertiary care referral center that provides health care services to a population of approximately 1.9 million people from the urban and rural area of Kavrepalanchowk, Sindhupalchowk, Dolakha, Sindhuli, Ramechhap, Bhaktapur, and other surrounding districts.

Participants

All the children visiting the department under 15 years of age were eligible for the study.

Exclusion criteria were set for convenience:

  1. A.

    Children with height outside of the Broselow tape’s limit (less than 46.5 cm and above 142.5 cm)

  2. B.

    Children with chronic disease and who were under medications for prolonged duration; outcome may be compromised if readings of the abovementioned categories were taken into account as the height and weight may be different from the general population.

  3. C.

    Children with medical emergencies who require immediate medical intervention; it would have inappropriately delayed in receiving the emergency care during busy hours of the emergency department.

The Broselow pediatric emergency tape (Armstrong Medical Industries, Lincolnshire, IL, USA, 2017 edition) available in the ED of the hospital was used for this study purpose. The Broselow tape has 9 colored zones. For the study purpose, the color zones were further grouped into 3 groups with 3 color zones into each group, viz, group 1 (< 10 kg)—grey, pink, and red; group 2 (10–18 kg)—purple, yellow, and white; and group 3 (> 18 kg)—blue, orange, and green. A sample of 96 participants was required in each group to detect a 10% difference between actual weight and the Broselow estimated weight with an aggregate sample size of 288 using the sample calculation for standard normal deviate assuming 50% as the proportion of the target population.

Demographic data was collected from the guardian using a pre-designed questionnaire after informed consent was obtained. The weight of the candidates was measured by standard calibrated weighing machine to the nearest 0.1 kg using an electronic weighing machine. The shoes and heavy clothing were removed before measuring the weight. Another investigator was blinded to the age, measured the weight of the child, and recorded the estimated weight of the child using the Broselow tape in supine position measured from head to heel. Using the age provided by the caregiver, the child’s weight was calculated using the updated APLS formulae—weight (kg): infant 0–12 months = (0.5 × age in months) + 4; children aged 1–5 years = 2 × age (years) + 8; 6–12 years = 3 × age (years) + 7.

The size of the internal diameter (ID) of the ET tube was calculated using Pediatric Advanced Life Support (PALS) recommendations: 3.5-mm ID uncuffed ET tube for infants up to 1 year of age and 4.0-mm ID uncuffed ET tube for children between 1 and 2 years of age. After age of 2, by the following formula: age/4 + 4 for uncuffed and age/4 + 3.5 for cuffed ET tube [6]. The size of the calculated ET tube was rounded off reducing to the nearest 0.5 mm ID. The calculated size was compared with the size recommended according to the corresponding Broselow zone. The calculated dose of adrenaline (0.01 mg/kg 1:10,000) was compared with the estimated dose according to the Broselow tape. ET tubes with an ID 0.5 mm smaller or larger than the estimated size are recommended to be made available during the intubation attempt. Therefore, an acceptable ET tube size was defined as the ET tube size ranging within 0.5 mm size of the calculated size. The acceptable dose of adrenaline was defined as a drug dose within the range of ± 20% from the recommended dose [9]. Any values beyond these recommendations were considered to be an error.

Data analysis

The categorical variables were presented by frequency and percentage, and the continuous variables by mean with standard deviation and range. The percentage of agreement and over- or underestimation within the one/two color zone of the estimated color zone according to the Broselow tape with the actual weight was calculated. The relationship between the actual body weight and the estimated body weight was analyzed by Pearson’s correlation test. The Bland-Altman analysis was performed to evaluate the measures of bias, precision, and accuracy [23]. The mean percentage error (MPE = mean of the sum of [estimated weight − actual weight/actual weight] × 100) limits of agreement (LOA) was used to evaluate the bias and precision, whereas the percentage of weight estimations within 10% and 20% of actual weight (PW10 and PW20) was calculated to evaluate the accuracy of the Broselow tape and the updated APLS formula. All the analysis was carried out using SPSS version 21. A p value less than 0.05 was considered significant.

Results

This study included 315 children with male to female ratio of 0.63:1. They were divided into 3 groups according to their estimated weight by the Broselow tape into < 10 (gray, pink, and red color code), 10–18 (purple, yellow, and white color code), and > 18 kg (blue, orange, and green color code) which comprised 100, 113, and 102 children, respectively. The demographic and descriptive details of the enrolled children are illustrated in Table 1.

Table 1 Demographics of the children

The agreement of the estimated color zone according to the Broselow tape with the actual weight is demonstrated in Table 2. A total agreement was seen in the gray color zone, and the convergence gradually decreased with increasing length and weight of the children (p = 0.01).

Table 2 Agreement between estimated color zone by the Broselow tape and the actual weight

The relationship between the actual body weight and the estimated body weight was analyzed which demonstrated a positive relationship (Fig. 1) (correlation (r = 0.970, p = 0.01) and accuracy (r2 = 0.941)). The crowding of data points is noticed below 10 kg and starts to scatter more as the weight of the child increases. This indicates that the accuracy of estimated weight with the Broselow tape decreases with increasing weight of children.

Fig. 1
figure 1

Correlation between the measured weight and the Broselow tape estimated weight. Correlation (r = 0.970, p = 0.01) and accuracy (r2 = 0.941)

The measures of bias and accuracy for the Broselow tape and APLS formula are shown in Table 3. The mean percentage error (MPE) with the Broselow tape was lowest for the < 10 kg (3.11%) and gradually increased in the 10–18 kg (10.57%) and > 18 kg (13.55%) groups. The Broselow tape demonstrated the similar bias of weight estimation compared to the APLS formula in the first two groups, except in the > 18 kg group, where the bias was more in APLS formula weight estimation group (13.55 versus 32.60). The predicted weight within the PW10 was highest among the < 10 kg group for both the Broselow tape and APLS formula (68.7% and 43.3%, respectively) and lowest among the > 18 kg group (35.3% and 20%, respectively) (Fig. 2).

Table 3 Measures of bias, precision, and accuracy for the Broselow tape and APLS formula for weight estimation
Fig. 2
figure 2

Comparison of PW 10% between weights estimated by the Broselow tape and modified APLS formula. Comparison of percent within 10% (PW10) of actual weight measured by weighing machine to that estimated by the Broselow tape and updated APLS formula

A Bland-Altman plot was drawn to evaluate the agreement between the actual weight and the MPE (Fig. 3). The upper and lower Bland-Altman limits of agreement were 6.18857 and − 3.32331, respectively, with a confidence interval of 95%. The plot demonstrates that the difference was scattered as the weight of the children increased; therefore, the precision of the tape was relatively high in the lower length zones as compared to the higher length zones. The Broselow tape estimated ET tube size was within the PALS recommendation and acceptable limit of within 0.5 size of recommended size in 88.76%, 94.9% in weight less than 10 kg, and 93.9% in weight 10–18 kg. Adrenaline drug dose estimation in weight group of < 18 kg is in accordance with actual weight-based calculation of drug dosing, showing 75.8% and 74.6% accuracy for weight group < 10 kg and 10–18 kg, respectively, which fails to predict the dose accurately as the weight increases, showing 64.7% for weight group of > 18 kg with acceptability limit of + 20% (Table 4).

Fig. 3
figure 3

The Bland-Altman plot of differences between actual weight and estimated weight by the Broselow tape. EW, estimated weight by the Broselow tape; AW, actual weight measured by weighing machine; upper limit (UL), 6.18857; lower limit (LL), − 3.32331; mean, 1.43263; confidence interval (CI), 95%

Table 4 ET tube size and adrenaline dose accuracy for estimated weight according to the Broselow tape

Discussion

This study explored the bias, precision, and accuracy of the Broselow tape in the prediction of the weight of children visiting the outpatient pediatric department and vaccination unit of a tertiary care hospital which serves the population from urban and rural areas of Nepal. The agreement of the estimated color zone according to the Broselow tape with the actual weight was closest in group 1 (< 10 kg) and gradually decreased with increasing length and weight of the children. The relationship between the actual body weight and the estimated body weight was analyzed which indicated that the accuracy of estimated weight with the Broselow tape decreases with increasing weight of children.

For pediatric resuscitation, precise weight estimation is critical as the emergency drugs, equipment, and shock dosage are based on the weight of the individual child. We used the 2017 version of the Broselow tape for this study which was available in the ED. A reference at each color zone on the tape suggests equipment sizes to perform emergency resuscitation and pre-calculated medication dosages. Pre-calculated doses in milligrams, as well as all doses in pre-calculated ml in the 2017 version, remove the need for further calculations and may decrease the error related to the calculation. The Broselow cart with 9 color-specific drawers which matched the colors on the Broselow tape was constructed locally and utilized in our ED for emergency pediatric care. With this method, the drug dosage and equipment sizing could easily and quickly be determined and accessed. However, it was a concern of inaccuracy in LMIC due to underweight, lack of research in the field. Moreover, it is a concern that the newer version of the tape is more inaccurate than older versions of the tape in low- and middle-income countries because its shifted weight zones due to raised concerns regarding underestimations of drug dosages secondary to obesity in high-income country growth data [24]. Using newer edition of the tape presumably creates a greater error in estimating the weight of our populations compared to prior edition as it has been modified to minimize weight underestimation in obese children in the USA.

There are studies done in various parts of the world providing the adaptability and efficiency of the Broselow tape in measuring the weight of children [10, 11, 25] while other studies had contradictory conclusions [13,14,15,16,17, 22]. Studies from Kenya [10] and South Africa [10, 25] showed that the Broselow tape estimated weights varied only minimally from the actual weight. Clark et al. discussed their concern about using the Broselow tape for weight estimation among children with malnutrition in low-income countries like Sudan. They concluded that both the Broselow tape and age-based formulas were markedly inaccurate in both the non-malnourished and malnourished African children. Moreover, it was unacceptable in malnourished children and they suggested further studies to explore appropriate methods of weight and dose estimation for children from regions with a high prevalence of malnutrition.

A study done in India also showed similar results to ours that the results were favorable in the children < 10 kg group and 10–18 kg group, and they also had the opinion that as pediatric weight increases, the reliability of the tape significantly plummeted [17]. The Broselow tape overestimated weight by > 10% in the majority of Indian children, and they suggested to apply 10% weight correction factor to the Broselow estimated weight [18, 19]. Sinha et al. [26] compared the actual weight with the Broselow estimated weight in pediatric trauma patients and concluded that the bias was greatest in the highest weight category.

The measures of bias and accuracy for the Broselow tape and APLS formula were similar for the < 10 kg and 10–18 kg group. While comparing the estimated Broselow weight against APLS formula, the former met better accuracy especially in children > 18 kg. Another study also demonstrated that weights estimated using the Broselow tape correlated better with actual weights than those calculated using the APLS and updated APLS formulae [19].

A single study which was done in Nepal [22] compared the measured weight with the estimated weight of Broselow, PAWPER XL [27], and Mercy tapes [28]. The study concluded that PAWPER XL tape provides the most accurate weight estimation of children in Nepal followed by the Broselow tape. The accuracy remained high for children weighing over 20 kg, unlike the other two methods. However, the PAWPER tape is not easily accessible in resource-limited settings. When we compare the results of our study with that of this study, the accuracy was lower for the Broselow tape in all weight groups, PW10 (49.8 versus 63.2) and PW20 (81.2 versus 91.7). This could be due to the different patient characteristics and use of the newer version of tape in our study.

The overestimation of equipment size and the drug doses may lead to serious side effects and delay in resuscitation. A study conducted in India showed the accuracy of the Broselow tape in estimating the ET tube size was 50.3% in children less than 6.5 years of age and weight of less than 15 kg [29]. We tested the equipment size (ET tube) and drug dose (adrenaline) to check for the accuracy of their estimation using the Broselow tape. The ET tube size and Adrenaline dose estimation by using Broselow Tape was within acceptable limit for weight group <10 kg and 10-18 kg.

Based on the findings of this study, we can demonstrate that the Broselow tape can be safely used in children weighing less than 18 kg and should be avoided in children weighing more than 18 kg. This finding contradicts our research hypothesis; the Broselow tape overestimates the weight, equipment size, and drug doses in pediatric age group; and drug overdose would affect the children with smaller weight. However, the study showed the Broselow tape to be more accurate in estimating the weight, equipment size, and drug doses in children with weight less than 18 kg.

Limitation

The study was conducted at a single hospital and cannot be generalized to the whole Nepalese population. Only the children attending the hospital are included in the study, not the general population which may not fulfill the true purpose of the study. Our study compared only limited drug dose and equipment size, but the Broselow tape recommends multiple other drug doses and equipment sizes. Our study could not provide recommendations for estimating weight of children in age of more than 18 kg, so further study is required in the field.

Conclusions

The Broselow tape is more reliable in children of the < 10 kg and 10–18 kg groups. However, as pediatric weight is increased, predictive reliability is also decreased, especially in those weighing > 18 kg. However, the Broselow tape was found to be superior to updated APLS formula in estimating the weight of child in all weight groups.

Thus, medical errors will be reduced by the implementation of the Broselow tape in Nepal to estimate pediatric weight accurately in the emergency setting for children of weight group < 18 kg, but an alternative method for weight estimation is suggested for weight group > 18 kg in Nepal.

Similarly, the estimation of ET tube size and adrenaline drug dose by using the Broselow tape is more accurate and reliable in children of < 18 kg and its predictive reliability decreases as weight increases. Thus, age-based PALS formula is recommended to be used for estimating ET tube size and weight-based formula for adrenaline dose in children above 18 kg weight group.

Availability of data and materials

The datasets used and/or analyzed during the current study are available from the corresponding author on reasonable request.

Abbreviations

APLS:

Advanced pediatric life support

ED:

Emergency department

ET tube:

Endotracheal tube

ID:

Internal diameter

PALS:

Pediatric Advanced Life Support

References

  1. Wong ICK, Ghaleb MA, Franklin BD, Barber N. Incidence and nature of dosing errors in paediatric medications: a systematic review. Drug Saf. 2004;27(9):661–70. https://doi.org/10.2165/00002018-200427090-00004.

    Article  PubMed  Google Scholar 

  2. Selbst SM, Fein JA, Osterhoudt K, Wayne HO. Medication errors in a pediatric emergency department. Pediatr Emerg Care. 1999;15:1–4. https://doi.org/10.1097/00006565-199902000-00001.

    Article  CAS  PubMed  Google Scholar 

  3. Wells M, Goldstein LN, Bentley A. It is time to abandon age-based emergency weight estimation in children! A failed validation of 20 different age-based formulas. Resuscitation. 2017;116:73–83. https://doi.org/10.1016/j.resuscitation.2017.05.018.

    Article  PubMed  Google Scholar 

  4. American College of Surgeons Committee on Trauma. In: ATLS Advanced Trauma Life Support: student course manual. N. Saint Clair Street, Chicago, Illinois: American College of Surgeons; 2018;10:190.

  5. Lubitz DS, Seidel JS, Chameides L, Luten RC, Zaritsky AL, Campbell FW. A rapid method for estimating weight and resuscitation drug dosages from length in the pediatric age group. Ann Emerg Med. 1988;17:576–81. https://doi.org/10.1016/s0196-0644(88)80396-2.

    Article  CAS  PubMed  Google Scholar 

  6. Kleinman ME, Chameides L, Schexnayder SM, Samson RA, Hazinski MF, Atkins DL, et al. Part 14: Pediatric Advanced Life Support: 2010 American Heart Association guidelines for cardiopulmonary resuscitation and emergency cardiovascular care. Circulation. 2010;122:S876–908. https://doi.org/10.1161/circulationaha.110.971101.

    Article  PubMed  Google Scholar 

  7. Agarwal S. Comparing the utility of a standard pediatric resuscitation cart with a pediatric resuscitation cart based on the Broselow tape: a randomized, controlled, crossover trial involving simulated resuscitation scenarios. Pediatrics. 2005;116:e326–33. https://doi.org/10.1542/peds.2005-0320.

    Article  PubMed  Google Scholar 

  8. Shah AN, Frush K, Luo X, Wears RL. Effect of an intervention standardization system on pediatric dosing and equipment size determination: a crossover trial involving simulated resuscitation events. Arch Pediatr Adolesc Med. 2003;157(3):229–36 Available from: https://www.ncbi.nlm.nih.gov/pubmed/12622671.

    Article  Google Scholar 

  9. Kaji AH, Gausche-Hill M, Conrad H, Young KD, Koenig WJ, Dorsey E, et al. Emergency medical services system changes reduce pediatric epinephrine dosing errors in the prehospital setting. Pediatrics. 2006 Oct;118(4):1493–500. https://doi.org/10.1542/peds.2006-0854.

    Article  PubMed  Google Scholar 

  10. House DR, Ngetich E, Vreeman RC, Rusyniak DE. Estimating the weight of children in Kenya: do the Broselow tape and age-based formulas measure up? Ann Emerg Med. 2013 Jan;61(1):1–8. https://doi.org/10.1016/j.annemergmed.2012.07.110.

    Article  PubMed  Google Scholar 

  11. AlHarbi TM, AlGarni A, AlGamdi F, Jawish M, Wani TA, Abu-Shaheen AK. The accuracy of Broselow tape weight estimate among pediatric population. Biomed Res Int. 2016;2016:1–8. https://doi.org/10.1155/2016/7042947.

    Article  Google Scholar 

  12. Al-Busaidi A, Jeyaseelan L, Al-Barwani H. The accuracy of the Broselow™ pediatric emergency tape for weight estimation in an Omani paediatric population. Sultan Qaboos Univ Med J. 2017:e191–5. https://doi.org/10.18295/squmj.2016.17.02.009.

  13. O’Leary F, John-Denny B, McGarvey K, Hann A, Pegiazoglou I, Peat J. Estimating the weight of ethnically diverse children attending an Australian emergency department: a prospective, blinded, comparison of age-based and length-based tools including Mercy, PAWPER and Broselow. Arch Dis Child. 2017;102:46–52. https://doi.org/10.1136/archdischild-2016-310917.

    Article  PubMed  Google Scholar 

  14. Wells M, Goldstein LN, Bentley A, Basnett S, Monteith I. The accuracy of the Broselow tape as a weight estimation tool and a drug-dosing guide - a systematic review and meta-analysis. Theatr Res Int. 2017;121:9–33. https://doi.org/10.1016/j.resuscitation.2017.09.026.

    Article  Google Scholar 

  15. Clark MC, Lewis RJ, Fleischman RJ, Ogunniyi AA, Patel DS, Donaldson RI. Accuracy of the Broselow tape in South Sudan, “the hungriest place on Earth”. Acad Emerg Med. 2016;23:21–8. https://doi.org/10.1111/acem.12854.

    Article  PubMed  Google Scholar 

  16. Al-Mendalawi M. The accuracy of Broselow pediatric emergency tape in estimating body weight of pediatric patients. Saudi Medical Journal. 2017;38:1158. https://doi.org/10.15537/smj.2017.11.21033.

    Article  PubMed  PubMed Central  Google Scholar 

  17. Mishra DG. A correlation analysis of Broselow pediatric emergency tape-determined pediatric weight with actual pediatric weight in India. World J Emerg Med. 2016;7:40. https://doi.org/10.5847/wjem.j.1920-8642.2016.01.007.

    Article  PubMed  PubMed Central  Google Scholar 

  18. Ramarajan N, Krishnamoorthi R, Strehlow M, Quinn J, Mahadevan SV. Internationalizing the Broselow tape: how reliable is weight estimation in Indian children. Acad Emerg Med. 2008;15(5):431–6. https://doi.org/10.1111/j.1553-2712.2008.00081.x.

    Article  PubMed  Google Scholar 

  19. Shah V, Bavdekar SB. Validity of Broselow tape for estimating weight of Indian children. Indian J Med Res. 2017 Mar;145(3):339–46. https://doi.org/10.4103/ijmr.IJMR_837_14.

    Article  PubMed  PubMed Central  Google Scholar 

  20. Nutrition [Internet]. UNICEF Nepal. [cited 2019 Jul 16]. Available from: https://www.unicef.org/nepal/nutrition

  21. Asskaryar F, Shankar R. An Indian pediatric emergency weight estimation tool: prospective adjustment of the Broselow tape. Int J Emerg Med. 2015;8(1):78. https://doi.org/10.1186/s12245-015-0078-z.

    Article  PubMed  Google Scholar 

  22. Shrestha K. Estimating the weight of children in Nepal by Broselow, PAWPER XL and Mercy method. World J Emerg Med. 2018;9:276. https://doi.org/10.5847/wjem.j.1920-8642.2018.04.007.

    Article  PubMed  PubMed Central  Google Scholar 

  23. Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1(8476):307–10 Available from: https://www.ncbi.nlm.nih.gov/pubmed/2868172.

    Article  CAS  Google Scholar 

  24. Ralston ME, Myatt MA. Weight estimation for children aged 6 to 59 months in limited-resource settings: a proposal for a tape using height and mid-upper arm circumference. PLoS One. 2018;13(6):e0197769. https://doi.org/10.1371/journal.pone.0197769.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  25. Geduld H, Hodkinson PW, Wallis LA. Validation of weight estimation by age and length based methods in the Western Cape, South Africa population. Emerg Med J. 2011 Oct;28(10):856–60. https://doi.org/10.1136/emj.2010.098640.

    Article  PubMed  Google Scholar 

  26. Sinha M, Lezine MW, Frechette A, Foster KN. Weighing the pediatric patient during trauma resuscitation and its concordance with estimated weight using Broselow Luten Emergency Tape. Pediatr Emerg Care. 2012 Jun;28(6):544–7. https://doi.org/10.1097/PEC.0b013e318258ac2e.

    Article  PubMed  Google Scholar 

  27. Wells M, Coovadia A, Kramer E, Goldstein L. The PAWPER tape: a new concept tape-based device that increases the accuracy of weight estimation in children through the inclusion of a modifier based on body habitus. Theatr Res Int. 2013 Feb;84(2):227–32. https://doi.org/10.1016/j.resuscitation.2012.05.028.

    Article  Google Scholar 

  28. Abdel-Rahman SM, Paul IM, James LP, Lewandowski A. Evaluation of the Mercy TAPE: performance against the standard for pediatric weight estimation. Ann Emerg Med. 2013;62:332–9.e6. https://doi.org/10.1016/j.annemergmed.2013.02.021.

    Article  PubMed  PubMed Central  Google Scholar 

  29. Subramanian S, Nishtala M, Ramavakoda CY, Kothari G. Predicting endotracheal tube size from length: evaluation of the Broselow tape in Indian children. J Anaesthesiol Clin Pharmacol. 2018;34(1):73–7. https://doi.org/10.4103/joacp.JOACP_317_16.

    Article  PubMed  PubMed Central  Google Scholar 

Download references

Acknowledgements

The authors thank the Department of Pediatrics, Dhulikhel Hospital, Kathmandu University Teaching Hospital, Kavrepalanchowk, Nepal.

Funding

No funding was received for the study.

Author information

Authors and Affiliations

Authors

Contributions

PK, AJ, and KG prepared the research proposal, collected the data, analyzed the data, and prepared the manuscript. AS and RS helped in preparing the research proposal, analyzed the data, and prepared the manuscript. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Pukar K.C..

Ethics declarations

Ethics approval and consent to participate

Ethical approval was obtained from the Kathmandu University School of Medical Sciences, Institutional Review Committee (KUSMS-IRC). Informed consent to participate in the study was obtained from the children’s parent or guardian.

Consent for publication

Not applicable

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

K.C., P., Jha, A., Ghimire, K. et al. Accuracy of Broselow tape in estimating the weight of the child for management of pediatric emergencies in Nepalese population. Int J Emerg Med 13, 9 (2020). https://doi.org/10.1186/s12245-020-0269-0

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12245-020-0269-0

Keywords