The use of dual oxygen concentrator system for mechanical ventilation during COVID-19 pandemic in Sabah, Malaysia

Abstract

Sabah in Malaysian Borneo is among the Malaysian states which reported a high number of detected COVID-19 cases during the current pandemic. Due to geographical challenges and limited resources, clinicians developed novel strategies for managing patients. The use of a dual oxygen concentrator system for mechanical ventilation is one of the innovations developed by retrieval team members from the Emergency Department (ED) of the Sabah Women and Children’s Hospital. Due to conditions requiring isolation of patients suspected of or positive for COVID-19, high-risk patients were treated in an ED extension area that lacked central wall oxygen. Direct access to oxygen tanks became the only viable option, but ensuring a continuous supply was laborious. The novel setup described within this paper has been used on intubated patients in the ED extension area with moderate to high ventilator settings successfully. This simple setup, designed to meet the limited resources within a pandemic environment, needed only a turbine-driven ventilator, two oxygen concentrators, a 3-way connector, and three oxygen tubing. The application of this setup could potentially save more critically ill patients who are being managed in resource-limited conditions such as in smaller district hospitals or out in the field.

Introduction

Sabah is a state in Malaysia located on the northern portion of Borneo. A major part of Sabah is surrounded by jungle and mountains which often pose a problem when it comes to transferring and treating patients in rural settings. All hospitals in Sabah started to prepare for the coronavirus disease 2019 (COVID-19) pandemic when the first case was detected on 17 March 2020 in the district of Beaufort. Since then, numbers had increased and hospitals in Sabah began to prepare to treat critically ill patients. Sabah Women and Children’s Hospital (SWACH) started preparation with limited resources, preparing to treat patients in the hospital, while the SWACH retrieval team was placed on standby for the transferring of critically ill patients.

Major hospitals in Sabah have constructed a separate area for treatment and resuscitation of suspected COVID-19 patients using disaster tents and shelters in an effort to segregate high- and low-risk patients arriving at the emergency department (ED). Patients are triaged according to their risk evaluation and if deemed to have a high risk of infection, are directed into the ED extension area [1, 2]. The ED extension was separated into different zones defined as non-critical, semi-critical, and critical. In the extension area, oxygen supply posed a great challenge as there was no central oxygen supply, so teams resorted to oxygen tanks and oxygen concentrators for supply.

For patient transfers, an average built patient with a body weight of 70 kg, with ventilator settings of tidal volume 420 mL, positive end-expiratory pressure (PEEP) 5mmHg, respiratory rate 12/min, and a fraction of inspired oxygen (FiO2) 0.5, at least 2 size E oxygen tanks are needed for a 2-h journey. This is equivalent to 16 kg of oxygen [3]. Transfer of patients from district to tertiary hospitals in Sabah is challenging due to the long distances between these hospitals. Our hospital retrieval teams are familiar with transferring ventilated patients using portable battery-operated oxygen concentrators when patients require low ventilator settings due to their ease of use compared to heavy and cumbersome oxygen tanks [4, 5]

Mechanically ventilating patients in everyday situations where there is adequate oxygen supply already requires highly technical skills from trained healthcare workers. Ventilating a patient in the setting of limited oxygen supply and resources would prove an even bigger challenge. At the beginning of the pandemic, we predicted prolonged patient stay in the ED extension area in view of intensive care units (ICU) having to come up with new standard operating procedures (SOP) which involved preparing new isolation bays with delays in receiving patients due to the meticulous steps of donning proper personal protective equipment (PPE). When ICUs are full with critically ill patients, subsequent patients requiring mechanical ventilation become the responsibility of the ED, beyond a duration that would be considered routine. Ventilated patients consume large volumes of oxygen and the changing and replacement of oxygen tanks in the separate treatment area was laborious with the multiple infection control protocols in place. In addition, the supply and refilling of oxygen tanks during the pandemic period were unreliable, and oxygen tanks would often run out during the night. Oxygen concentrators were a great substitute for tank oxygen supply albeit its limited applications as it was not able to supply high-pressure oxygen to our standard oxygen driven ventilators.

In an effort to reduce the use of oxygen tanks in the ED extension area, we paired our turbine-driven ventilator with oxygen concentrators for the ventilation of patients. We limited the single oxygen concentrator technique to patients who needed only low-minute volume settings. However, for patients on high ventilator settings with higher oxygen demand, we used a dual oxygen concentrator system allowing us to double the oxygen inflow to the turbine-driven ventilator. Each oxygen concentrator is only able to contribute a maximum of 5L/min of oxygen flow. Therefore, by connecting two oxygen concentrators using a 3-way connector, we were able to provide a maximum of 10 L/min of oxygen flow to the ventilator.

Our oxygen concentrator, Phillips EverFlo [6], is capable of generating 0.5–5 L/min of continuous flow oxygen. This device produces concentrated oxygen from normal room air for delivery to a patient requiring low-flow oxygen therapy. Oxygen from the air is concentrated using a molecular sieve and a pressure swing adsorption process. It does not have a built-in battery and therefore requires a continuous power supply [7]. The ventilator used was a Weinmann-Meduvent Standard [8], a turbine-driven ventilator capable of receiving low-flow oxygen to a maximum rating of 15 L/min. This ventilator has an oxygen sensor which detects the oxygen concentration being delivered to the patient, and therefore, this reading might vary from breath to breath according to the clinical condition and ventilator settings.

To set the FiO2, the minute ventilation administered was used as a guide [9]. According to the desired target oxygen graph provided by the ventilator manufacturer, a 14L/min of oxygen flow will be able to deliver FiO2 of 1.0 to the patient at administered minute volume (AMV) 15L/min body temperature and pressure, saturated (BTPS). At oxygen flow of 9L/min, FiO2 of 1.0 can be delivered if the AMV was 10L/min BTPS. By combining 2 oxygen concentrators, we estimate that we will be able to deliver 10L/min of oxygen to the ventilator with 10% subtracted for leaking. With this, we estimate that we will be able to safely ventilate a patient using the dual oxygen concentrator system up to a required AMV of 10L/min at FiO2 1.0 [9]. PEEP does not play a role in determining the oxygen concentration delivered [10]. For any patient with a higher ventilatory requirement, this system would not be used. Initial oxygen flow settings are based on the estimation above, but FiO2 levels shown on the ventilator will be watched closely and the oxygen flow will be adjusted to achieve the required FiO2 once the patient was connected to the ventilator. For clarity, the required FiO2 is adjusted by adjusting the flow rate of oxygen on the oxygen concentrator while the delivered FiO2 is monitored by looking at the FiO2 levels on the ventilator.

The 3-way connector used to connect oxygen tubing from the two oxygen concentrators was originally designed for the infusion of fluid [11]. While not optimal, its selection was based on limited resources and the lack of a more suitable alternative. The connector was modified to fit the oxygen tubing and act as a connection to direct the flow of both oxygen concentrators into the ventilator. Another two oxygen tubings were connected to the two other ends of the 3-way connector and oxygen concentrators respectively (Fig. 1). The connection system was then checked for leaks, and if present, adjusted or taped. One oxygen tubing [12] is cut at one end in order to attach to the 3-way connector, and subsequently, the other end will be connected to the ventilator.

Fig. 1
figure 1

Two oxygen concentrators connected to a low-flow turbine-driven ventilator using a three-way connector

An oxygen concentrator is a low-maintenance source for supplemental oxygen. In our ED, the oxygen concentrators are able to deliver a maximum of 5L/min of continuous flow oxygen. Continuous flow oxygen is straightforward compared to pulse oxygen. In its conventional usage, continuous flow oxygen concentrators deliver oxygen throughout the breath which leads to a significant fraction of delivered oxygen being lost to the ambient air and never reaching the patient’s lungs. Pulse dose concentrators aim to improve efficiency by limiting oxygen delivery only when patient inspiration can be detected by the trigger mechanism, delivering oxygen as a bolus or pulse early in the breath so that the fraction of oxygen exhaled from the anatomical dead space is reduced. Although usage of pulsed dosed oxygen is advantageous to reduce power consumption, this method of delivery is not suitable for use with this system as the trigger mechanism for pulse oxygen will not be activated when connected to a ventilator.

The main limitation of this particular oxygen concentrator is that a continuous power supply is needed as it does not come with a built-in battery. For transportation of patients, battery-operated oxygen concentrators are used. Although we were still able to achieve the required FiO2 with this system, we noticed minimal leaking at the connection between the three-way connector and the oxygen tubing as it was not designed to be used in this manner. To overcome this problem, an oxygen Y-connector [13] could be used for a better seal.

A single oxygen concentrator may be used with a turbine-driven ventilator, but this dual oxygen concentrator system is needed for patients on higher minute volume settings to achieve the desired FiO2 of 1.0 in certain patients. This system is suitable for district hospitals, private general practitioner offices, and also on the field to overcome limited oxygen supply. This setup should only be used with turbine ventilators equipped with real-time oxygen sensors due to the possible fluctuation of FiO2 with each patient’s breath.

The use of the dual oxygen concentrator system for mechanical ventilation is feasible as a temporary measure until the patient can be transferred to a proper critical care area like the ICU. It is extremely useful in a situation like the COVID-19 pandemic where oxygen supply was limited due to logistics and supply chain constraints. The method described above is a preliminary concept which requires further validation and should be used cautiously if needed.

Availability of data and materials

Data sharing is not applicable to this article as no datasets were generated or analyzed during the current study.

Abbreviations

COVID-19:

Coronavirus disease 2019

SWACH:

Sabah Women and Children’s Hospital

ED:

Emergency department

PEEP:

Positive end-expiratory pressure

FiO2 :

Fraction of inspired oxygen

ICU:

Intensive care units

SOP:

Standard operating procedures

PPE:

Personal protective equipment

AMV:

Administered minute volume

BTPS:

Body temperature and pressure, saturated

References

  1. CDC. sop-triage-prevent-transmission. https://www.cdc.gov/coronavirus/2019-ncov/hcp/non-us-settings/sop-triage-prevent-transmission.html. Publised April 16, 2020. Accessed 16 June 2020.

  2. KKM. garis-panduan-kkm. http://covid-19.moh.gov.my/garis-panduan/garis-panduan-kkm. Publishes June 16, 2020. Accessed 16 June 2020.

  3. Lutman D, Petros AJ. How many oxygen cylinders do you need to take on transport? A nomogram for cylinder size and duration. Emerg Med J. 2006;23(9):703–4. https://doi.org/10.1136/emj.2005.026724.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  4. Shrestha BM, Singh BB, Gautam MP, Chand MB. The oxygen concentrator is a suitable alternative to oxygen cylinders in Nepal. Can J Anesth. 2002;49(1):8–12. https://doi.org/10.1007/BF03020412.

    Article  PubMed  Google Scholar 

  5. Rybak M, Huffman LC, Nahouraii R, Loden J, Gonzalez M, Wilson R, et al. Ultraportable oxygen concentrator use in U.S. Army special operations forward area surgery: a proof of concept in multiple environments. Mil Med. 2017;182(1):e1649–52. https://doi.org/10.7205/milmed-d-16-00100.

    Article  PubMed  Google Scholar 

  6. Philips. EverFlo Home Oxygen System. https://www.philips.com.my/healthcare/product/HC0044000/everflo-home-oxygen-system/documentation. Accessed 16 June 2020.

  7. Philips. Ever Flo/Ever Flo Q user manual table of contents. Respironics; 2014:3-8. https://www.manualslib.com/manual/1423647/Philips-Everflo.html. Accessed 16 June 2020.

  8. Weinmann Medical Technology. Meduvent Standard. https://www.weinmann-emergency.com/products/medical-ventilators/meduvent-standard/. Accessed 16 June 2020.

  9. Weinmann Emergency. MEDUVENT Standard ventilator instructions for use, vol. 63: Weinmann; 2020. https://www.weinmann-emergency.com/download/. Accessed 16 June 2020

  10. Bordes J, d’Aranda E, Savoie PH, Montcriol A, Goutorbe P, Kaiser E. FiO2 delivered by a turbine portable ventilator with an oxygen concentrator in an Austere environment. J Emerg Med. 2014;47(3):306–12. https://doi.org/10.1016/j.jemermed.2014.04.033.

    Article  PubMed  Google Scholar 

  11. Braun. Discofix-3-way-stopcock. https://www.bbraun.com/en/products/b/discofix-3-way-stopcock.html. Accessed 16 June 2020.

  12. Intersurgical. Oxygen-tubing. https://www.intersurgical.com/products/oxygen-and-aerosol-therapy/oxygen-tubing. Accessed 16 June 2020.

  13. Mediplas. Product catalogue silantes product catalogue silantes. Mediplas Respiratory Product; 13. http://www.mediplas.com.my/pdf/Mediplas_Cat_2015.pdf. Accessed 16 June 2020.

Download references

Acknowledgements

The authors would like to thank the Director-General of Health Malaysia for the permission to publish this paper.

Funding

Not applicable.

Author information

Authors and Affiliations

Authors

Contributions

PKC, SEM, KKN, and AKMH conceived the presented idea. PKC, RNP, and MIH developed the theory and the usage protocol with relevant calculations and safety precautions. PKC, SEM, KKN, AKMH, and MIH performed the procedure on multiple patients and improved the techniques. All authors discussed the technique and contributed to the final manuscript. The author(s) read and approved the final manuscript.

Corresponding author

Correspondence to Phee Kheng Cheah.

Ethics declarations

Ethics approval and consent to participate

Not applicable.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Cheah, P.K., Steven, E.M., Ng, K.K. et al. The use of dual oxygen concentrator system for mechanical ventilation during COVID-19 pandemic in Sabah, Malaysia. Int J Emerg Med 14, 30 (2021). https://doi.org/10.1186/s12245-021-00354-9

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12245-021-00354-9